Form 1459 (01.13) ©SOR Inc.
Press the “Finish” button to complete the calibration.
NOTE: It may be desirable to save an archive of the device calibration when the
device has been calibrated to a satisfactory zero and span output. To do this, press
the “Save Report” button. You will be prompted for a directory to save the le.
Once the directory is chosen, it will automatically be recalled upon subsequent
saves.
After the “Finish” button is pressed,
the following confirmation will be
displayed to confirm the calibration
process was completed successfully.
Switch Type:
Single Point: Closed -> Opened