Form 1588 (09.14) ©SOR Inc. 1/24815DT Smart Differential Pressure TransmitterGeneral InstructionsThese instructions provide information for installati
10/24 Form 1588 (09.14) ©SOR Inc. HART CommandDescriptionCommand Data Command Reply DataByte Description Byte Description9(cont)Read device va
Form 1588 (09.14) ©SOR Inc. 11/24HART CommandDescriptionCommand Data Command Reply DataByte Description Byte Description11Read unique identifier assoc
12/24 Form 1588 (09.14) ©SOR Inc. HART CommandDescriptionCommand Data Command Reply DataByte Description Byte Description18Write tag, descript
Form 1588 (09.14) ©SOR Inc. 13/24HART CommandDescriptionCommand Data Command Reply DataByte Description Byte Description34Write PV damping value0-3Dam
14/24 Form 1588 (09.14) ©SOR Inc. HART CommandDescriptionCommand Data Command Reply DataByte Description Byte Description48(cont)Read addition
Form 1588 (09.14) ©SOR Inc. 15/24HART CommandDescriptionCommand Data Command Reply DataByte Description Byte Description53Write device variable units5
16/24 Form 1588 (09.14) ©SOR Inc. HART CommandDescriptionCommand Data Command Reply DataByte Description Byte Description132Device configurati
Form 1588 (09.14) ©SOR Inc. 17/24HART CommandDescriptionCommand Data Command Reply DataByte Description Byte Description134Device configuration0"
18/24 Form 1588 (09.14) ©SOR Inc. Modbus RegisterRegister Definition Read/Write"Data Type"Description0 Loop current/voltage R Float
Form 1588 (09.14) ©SOR Inc. 19/24Modbus RegisterRegister Definition Read/Write"Data Type"Description80 R UINT8 BSL firmware version84 R UIN
2/24 Form 1588 (09.14) ©SOR Inc. NAMEPLATEAREAEXTERNALGROUNDPROVISIONSET SCREWANNUNCIATIONPATH2nd Step: Make the Electrical Connection The el
20/24 Form 1588 (09.14) ©SOR Inc. 34.11.34146.25-3/4158.96-1/427.91.1055.92.20ELECTRICALCABLEELECTRICALCONNECTION1/2 NPTM28.61-1/8 FLATSZEROMA
Form 1588 (09.14) ©SOR Inc. 21/24ATEX Marking DetailsSerial Number (First Two Numbers Indicate Year of Manufacture)Manufacturer’s Registered Trademark
22/24 Form 1588 (09.14) ©SOR Inc. ATEX Marking DetailsFor ATEX Certi ed Models EC Declaration of Conformity 14685 West 105th St
Form 1588 (09.14) ©SOR Inc. 23/24
24/24 Form 1588 (09.14) ©SOR Inc. 14685 West 105th Street, Lenexa, KS 66215 913-888-2630 800-676-6794 USA Fax 913-888-0767Register
Form 1588 (09.14) ©SOR Inc. 3/24NAMEPLATE AREASET SCREWANNUNCIATION PATHEXTERNAL GROUNDPROVISIONElectrical Termination - 4-20 mA OutputDrawing 0091546
4/24 Form 1588 (09.14) ©SOR Inc. THIS DRAWING IS THNO USE WHATSOEVER OHEREON, NOR REPRODUMADE WITHOUT THE EXPTOLERANCES:(UNLESS OTHERWISE FRAC
Form 1588 (09.14) ©SOR Inc. 5/24 Windowed, Normally-ClosedThe 815DT switch Offcan be configured Windowed, Normally-Opennine different ways: Wi
6/24 Form 1588 (09.14) ©SOR Inc. In this configuration, the switch output will be closed for pressures less than the selected setpoint. The s
Form 1588 (09.14) ©SOR Inc. 7/24RL > 250The table below contains a short description of the four 815DT device variables.Device Definition TableVar
8/24 Form 1588 (09.14) ©SOR Inc. 1-5 VDC OutputHART CommandsThe following pages summarize all HART commands currently supported by the 815DT.H
Form 1588 (09.14) ©SOR Inc. 9/24HART CommandDescriptionCommand Data Command Reply DataByte Description Byte Description0(cont)Read unique identifierRe
Kommentare zu diesen Handbüchern